Analisis Penghitungan Dosis Serap Terapi Rotasi Dengan Metode Tissue Phantom Ratio (TPR) Pada Linear Accelerator (LINAC) 6 MV
Intisari
Telah dilakukan penelitian tentang perhitungan dosis serap terapi rotasi
dengan metode Tissue Phantom Ratio (TPR) pada Linear Accelerator (LINAC).
Tujuan dari penelitian ini yaitu untuk mengetahui perbandingan dosis serap terapi
rotasi berdasarkan perhitungan dan pengukuran. Penelitian dilakukan pada solid
water phantom dengan SAD 100 cm. Penelitian dilakukan pada luas lapangan
persegi 5 cm x 5 cm ,7 cm x 7 cm dan 10 cm x 10 cm serta pada luas lapangan
ekivalen 10 cm x 10 cm yaitu 8 cm x 13,3 cm dan 9 cm x 11,5 cm. Dari hasil
penelitian didapat nilai persentase selisih antara dosis perhitungan dan pengukuran
pada masing-masing luas lapangan. Luas lapangan yang memiliki persentase di
bawah nilai toleransi yaitu 7 cm x 7 cm dengan nilai persentase selisih sebesar 0,46
%. Untuk luas lapangan 5 cm x 5 cm memiliki persentase selisih sebesar 2,65 %
sedangkan luas lapangan 10 cm x 10 cm dan lapangan ekivalennya memiliki
persentase selisih yang tinggi yaitu lebih dari 3%.
Kata kunci : terapi rotasi, TPR, lapangan ekivalen
Abstract
This Research has been conducted for the absorbed dose calculation in
rotational treatment with Tissue Phantom Ratio (TPR) method on 6 MV Linear
Accelerator (LINAC). The purpose of this study to determine the absorbed dose
comparisons on rotational therapy based on calculations and measurements. The
study was conducted in a solid water phantom with 100 cm SAD. The study was
conducted on a square field area of 5 cm x 5 cm, 7 cm x 7 cm and 10 cm x 10 cm as
well as on equivalent field area of 10 cm x 10 cm, consist of 8 cm x 13.3 cm and
9 cm x 11.5 cm area. The result is percentage of difference between the value of
measurements and calculations dose on each field area. The field area which have
percentage below the tolerance value is a 7 cm x 7 cm with the percentage of
difference of 0.46%. For the field area of 5 cm x 5 cm has a percentage of difference
of 2.65% , while the area of 10 cm x10 cm and equivalent field area the percentage
of difference were high at more than 3%.
Keywords: rotational therapy, TPR, equivalent field