• 024-7474754 (current)
  • fsm@undip.ac.id
    ``````
  • Visitor
  • Unduh
  • Pilih Bahasa :
    Bahasa Arab Bahasa Bengal Bahasa Brazil Portugis Bahasa Inggris Bahasa Spanyol Bahasa Jerman Bahasa Indonesia Bahasa Jepang Bahasa Melayu Bahasa Persia Bahasa Rusia Bahasa Thailand Bahasa Turki Bahasa Urdu

Perpustakaan Fakultas Sains Matematika

Universitas Diponegoro

  • Berita
  • Profil
    Profil Singkat Struktur Organisasi Sambutan Kepala Perpustakaan Pustakawan Kontak
  • E-Resources
    Indonesiana
    • Batavia Digital
    • Candi di Indonesia
    • Dokumentasi Perfilman Indonesia
    • Dokumentasi Sastra Indonesia
    • Kepustakaan Presiden RI
    • Keraton Nusantara
    • Pernaskahan Nusantara
    • Pusaka Indonesia
    • Kepustakaan Tokoh Pahlawan P. Diponegoro
    • Kepustakaan Tokoh Pahlawan Jenderal Soedirman
    • Kepustakaan Tokoh Perfilman
    • Khasanah Pustaka Nusantara
    • Perpuspedia
    • Literasi Kanker Indonesia
    Pencarian
    • Indonesia One Search
    • DOAJ
    • Google Scholar
    • Scopus
    • JDIH
    • Bibliografi Nasional Indonesia
    • Katalog Induk Nasional
    • KINK Kemenkes
    • R2KN Kemenkes
    Alat & Sumber Belajar
    • Mendeley
    • Perpustakaan Digital
    • TED
    • Google Experiments
    • Sumber Belajar Kemdikbud
    • Ebook
    Pustaka Kami
    • Jurnal Nasional
    • Jurnal Internasional
    • Jurnal Dilanggan
    • Prosiding
    • Modul Bahan Ajar
  • Panduan
    Perpustakaan Layanan FAQ Penulisan Penelusuran Informasi Akses Internet Koleksi
  • Layanan
    Daftar Anggota Online Sirkulasi Referensi dan Serial Bebas Pinjam Kartu Sakti
  • Area Anggota

Pencarian berdasarkan :

SEMUA Pengarang Subjek ISBN/ISSN Pencarian Spesifik

Pencarian terakhir:

{{tmpObj[k].text}}
No image available for this title

Text

Tingkat Kinerja, Dosis Radiasi Dan Kualitas Citra Computed Tomography Scanner (CT SCAN) Pada Mode Pemeriksaan Kepala

Samburi - Nama Orang;

ABSTRAK
Telah dilakukan penelitian pada CT single slice (SSCT), CT 16 – slice
dan CT 64-slice untuk menganalisis tingkat kinerja, dosis radiasi dan kualitas
citra. Proses penelitian dimulai dengan pengukuran tingkat kinerja dan dosis
radiasi menggunakan seperangkat multimeter dan fantom CTDI kepala. Proses
berikutnya adalah pengukuran kualitas citra menggunakan fantom Gammex ACR
464.
Hasil penelitian menunjukkan bahwa simpangan maksimum akurasi kV
dari SSCT berkisar 1,35-3,40%, CT 16-slice berkisar 0,91-3,88% dan CT 64-
slice berkisar 0,01-1,51%. HVL pada 120 kV semua tipe CT diatas batas
keberterimaan (3,8 mmAl). Akurasi keluaran radiasi 120 kV/100 mAs dan
koefisien linearitas (CL) SSCT berkisar 21,52-30,67 mGy dan 0,017 - 0,299,
CT 16-slice berkisar 22,37-28,14 mGy dan 0,008 - 0,018 dan CT 64-slice
berkisar 12,27 mGy - 17,36 mGy dan 0,004 - 0,010. Simpangan baku pola sebaran
radiasi CTDI100 dari SSCT berkisar 5,04 – 6.98, 16-slice berkisar 2.78 – 5.46
dan 64-slice berkisar 1.86 - 2.00. Hasil pengukuran kualitas citra semua CT Scan,
respon linearitas berada pada kurva kemiringan (R) diatas nilai acuan (R ≤ 0.99).
Homogenitas akurasi CT number, keseragaman noise dan keseragaman CTpusat
dan CTtepi berada pada rentang keberterimaan (± 5 HU). Contrast to Noise Ratio
(CNR) berkisar antara 1,0 – 2,1 sesuai dengan kriteria keberterimaan (CNR ≥
1.0). Serta hasil evaluasi resolusi spasial berkisar antara 6 - 7 lp/cm sesuai dengan
kriteria keberterimaan (≥ 6 lp/cm).
Kata Kunci : SSCT, MSCT, Tingkat Kinerja, Dosis Radiasi, Kualitas Citra

ABSTRACT
Research on performance level, radiation dose and CT scan image quality
has been done. The process of measuring the level of performance and dose of
radiation using a set of multimeter and head CTDI fantom to obtain tube voltage
accuracy, HVL, linearity dose and CTDI100. Evaluation of image quality using
fantom Gammex ACR 464. The data obtained is normalized to obtain constant
image noise.
The results showed that the maximum deviation of kV accuracy from
SSCT ranged from 1.35-3.40%, CT 16-slice ranged from 0.91-3.88% and CT 64-
slice ranged from 0.01 to 1.51%. HVL at 120 kV all CT types above the
acceptability limit (3.8 mmAl). The accuracy of radiation output of 120
kV/100mAs and linearity coefficient (CL) of SSCT ranged 21.52 - 30.67 mGy
and 0.017 - 0.299, CT 16-slice ranged 22.37 - 28.14 mGy and 0.008 - 0.018 and
CT 64- slices range 12.27 - 17.36 mGy and 0.004 - 0.010. The standard deviation
of CTDI100 radiation pattern from SSCT ranged from 5.04 to 6.98, 16-slices
ranging from 2.78 to 5.46 and 64-slice ranging from 1.86 to 2.00. Result of image
quality measurement, linearity response show that slope curve (R) all CT Scan
above reference value (≤0,99). Result of image quality measurement of all CT
Scan, linearity response is on slope curve (R) above reference value (R ≤ 0.99).
The homogeneity of CT number accuracy, noise uniformity and uniformity of
CTcenter and CTedge are in the acceptance range (± 5 HU). Contrast to Noise Ratio
(CNR) ranges from 1.0 - 2.1 according to acceptance criteria (CNR ≥ 1.0). And
the result of evaluation of spatial resolution ranged from 6 - 7 lp/cm in accordance
with acceptance criteria (≥ 6 lp/cm).
Keywords: SSCT, MSCT, Performance Level, Radiation Dose, Image Quality.


Ketersediaan
035S2FIS035 S2FIS 17-ivPerpustakaan FSM Undip (Referensi)Tersedia
Informasi Detail
Judul Seri
MAGISTER FISIKA
No. Panggil
035 S2FIS 17-iv
Penerbit
: ., 2017
Deskripsi Fisik
-
Bahasa
Indonesia
ISBN/ISSN
-
Klasifikasi
1575
Tipe Isi
-
Tipe Media
-
Tipe Pembawa
-
Edisi
-
Subjek
organic chemistry
Info Detail Spesifik
-
Pernyataan Tanggungjawab
Samburi
Versi lain/terkait

Tidak tersedia versi lain

Lampiran Berkas
Komentar

Anda harus masuk sebelum memberikan komentar

Perpustakaan Fakultas Sains Matematika
  • Informasi
  • Layanan
  • Pustakawan
  • Area Anggota

Jl. Prof. Sudarto, Tembalang, Kec. Tembalang, Kota Semarang, Jawa Tengah 50275

Tentang Kami

As a complete Library Management System, SLiMS (Senayan Library Management System) has many features that will help libraries and librarians to do their job easily and quickly. Follow this link to show some features provided by SLiMS.

Cari

masukkan satu atau lebih kata kunci dari judul, pengarang, atau subjek

Donasi untuk SLiMS Kontribusi untuk SLiMS?

© 2026 — Perpustakaan Fakultas Sains Matematika

Ditenagai oleh SLiMS
Pilih subjek yang menarik bagi Anda
  • Karya Umum
  • Filsafat
  • Agama
  • Ilmu-ilmu Sosial
  • Bahasa
  • Ilmu-ilmu Murni
  • Ilmu-ilmu Terapan
  • Kesenian, Hiburan, dan Olahraga
  • Kesusastraan
  • Geografi dan Sejarah
Icons made by Freepik from www.flaticon.com
Pencarian Spesifik