Analisis Pengaruh Diameter Fantom Terhadap Computed Tomography Dose Index (CTDI) Dan Faktor Konversi (f) Untuk Size-Specific Dose Estimates (SSDE)"
INTISARI
Pendahuluan: Tujuan dari penelitian ini adalah untuk menetapkan faktor
konversi untuk menormalkan keluaran dosis dari CT Scan ( volumetric computed
tomography dose index, CTDIvol) ke dosis pasien (size-specific dose estimate,
SSDE) untuk berbagai diameter phantom dan tegangan tabung.
Bahan dan Metode: Dalam penelitian ini, in-house phantom silindris akrilik
dengan diameter fisik mulai dari 8 cm hingga 40 cm dikembangkan. Setiap
phantom memiliki lubang di tengah dan empat lubang di area periferal. Phantom
dipindai menggunakan Siemens Somatom Definition AS CT Scanner dengan 200
mA, ketebalan irisan 10 mm dan tegangan tabung yang berbeda, yaitu 80, 100,
120 dan 140 kVp. Dosis di setiap lubang dan setiap phantom diukur menggunakan
Sensor Raysafe X2 CT. Nilai SSDE weighted (SSDEw) dihitung dari lima lubang
di setiap pengukuran. Faktor konversi ukuran untuk CTDI phantom kepala dan
body dibuat dengan membagi SSDEw untuk berbagai ukuran dengan SSDEw
pada diameter setara air masing-masing 33,90 cm dan 16,95 cm.
Hasil: Ditemukan bahwa faktor konversi ukuran secara eksponensial menurun
dengan meningkatnya ukuran phantom. Juga ditemukan bahwa faktor konversi
ukuran dipengaruhi oleh tegangan tabung. Perbedaan faktor konversi ukuran
antara 80 kVp dan 140 kVp untuk pasien yang sangat kecil dan sangat gemuk
lebih dari 15%.
Kesimpulan: Perubahan tegangan tabung harus dipertimbangkan untuk estimasi
dosis yang lebih akurat, di samping ukuran pasien.
Kata kunci: CTDIvol, dosis keluaran CT, dosis pasien, size-specific dose estimate
(SSDE), faktor konversi ukuran.
ABSTRACT
Introduction: The purpose of this study was to establish the conversion factors to
normalise the output dose of CT (i.e. volumetric computed tomography dose
index, CTDIvol) to the patient dose (i.e. size-specific dose estimate, SSDE) for
various phantom diameters and tube voltages.
Material and Methods: In this study, in-house cylindrical acrylic phantoms with
physical diameters ranging from 8 cm to 40 cm were developed. Each phantom
had a hole in the centre and four holes in the peripheral areas. The phantoms were
scanned using a Siemens Somatom Definition AS CT Scanner with 200 mAs, 10
mm slice thickness and different tube voltages, i.e. 80, 100, 120 and 140 kVps.
Doses in every hole and every phantom were measured using a Raysafe X2 CT
Sensor. The weighted SSDE (SSDEw) values were calculated from five holes in
every measurement. The size-conversion factors for the body and the head CTDI
phantoms were established by dividing the SSDEw for various size with the
SSDEw at the the water-equivalent diameter of 33.90 cm and 16.95 cm,
respectively.
Results: It was found that the size-conversion factor exponentially decreases with
the increase of phantom size. It was also found that the size-conversion factor is
affected by the tube voltage. The different of size-conversion factor between 80
kVp and 140 kVp for very small and very obese patient was more than 15%.
Conclusion: A more accurate dose estimation, an impact of the tube voltage
should be considered, in addition the size of patient.
Keywords: CTDIvol, CT output dose, patient dose, size-specific dose estimate
(SSDE), size-conversion factor